Multi-function LCD with numerical value and 30-segment bar graph 9V Manganese battery (6F22) Dimensions 123mm (L) × 73mm (W) × 25mm (D) Weight Approx. 170g (including battery) 3. LCD Screen and Controls

: Used to calibrate the meter to zero in an uncharged area before starting measurements. HOLD Button

The FMX-002 uses two integrated LED guide lights to ensure the correct measuring distance.

Replace the battery when the low-capacity indicator appears. Standard battery life is approximately 30 hours. simco fmx002 manual full

For voltages below 1.49 kV, the display shows the exact value (e.g., 0.55 kV).

Slide open the battery compartment cover on the back of the device. Connect a fresh 9V battery to the snap terminals.

should be calibrated annually to maintain accuracy, typically done through a certified Simco-Ion service center. Multi-function LCD with numerical value and 30-segment bar

The face of the FMX-002 features an LCD screen and three multi-function buttons:

Energize the metal calibration plate to +1,000V DC using the reference power supply. Verify that the FMX-002 reads 1.00 (±10%).

If the sensor plate is contaminated with dust or oils, the readings will be inaccurate. Clean with an antistatic cleaner and a lint-free cloth. 4. Troubleshooting and Maintenance Potential Cause No display/weak reading Low battery Replace 9V battery. Erratic readings Poor grounding Check grounding cord connection. Reading stuck at 0 Sensor blocked Clean the front sensing plate. Calibration: The HOLD Button The FMX-002 uses two integrated LED

is designed for ease of use and high accuracy in identifying electrostatic charge buildup.

The Simco FMX-002 is a compact, pocket-sized fieldmeter designed to measure electrostatic voltages up to

Which would you prefer?

Finally, the manual addresses troubleshooting common issues. This section acts as a first-line defense against equipment downtime. It categorizes problems such as "no airflow," "ionization failure," or "balance error," providing logical flowcharts for diagnosis. By empowering technicians to resolve minor issues—such as a loose connection or a dirty filter—the manual reduces reliance on external service engineers and minimizes production interruptions.

The FMX-002 is a precision electronic instrument containing a sensor highly sensitive to mechanical shock and vibration. The official manual includes several crucial warnings:


TOP